Evaluation Of A Semi-Automated Indirect Immunofluorescence Assay Using Validated In-House Hep-2 Cell Slides As A Reference Standard For Antinuclear Antibody Detection

Authors

  • Muhammad Zain Arshad Department of Immunology, Armed Forces Institute of Pathology/National University of Medical Sciences (NUMS) Rawalpindi Pakistan
  • Muhammad Omair Riaz Department of Immunology, Armed Forces Institute of Pathology/National University of Medical Sciences (NUMS) Rawalpindi Pakistan
  • Muhammad Hussain Department of Immunology, Armed Forces Institute of Pathology/National University of Medical Sciences (NUMS) Rawalpindi Pakistan
  • Dawood Ahmad Department of Immunology, Armed Forces Institute of Pathology/National University of Medical Sciences (NUMS) Rawalpindi Pakistan
  • Ayesha Tanveer Department of Radiology, Combined Military Hospital Rawalpindi/National University of Medical Sciences (NUMS) Pakistan
  • Muhammad Aftab Hassan Department of Immunology, Armed Forces Institute of Pathology/National University of Medical Sciences (NUMS) Rawalpindi Pakistan

DOI:

https://doi.org/10.51253/pafmj.v76i6.10690

Keywords:

Antinuclear Antibodies, HEp-2 Cells, Indirect Immunofluorescence, Semi-automated Indirect Immunofluorescence Assay

Abstract

Objective: To evaluate the analytical performance of a semi-automated assay for detection of antinuclear antibodies by indirect immunofluorescence.

Study Design: Cross-sectional study

Place and Duration of Study: Department of Immunology, Armed Forces Institute of Pathology (AFIP), Rawalpindi, Pakistan, from Jan to Jun 2023

Methodology: A total of thirty-six serum samples were selected from patients with antinuclear antibody-positive systemic autoimmune rheumatic diseases (n = 26) and from healthy controls (n = 10). Samples were applied on in-house HEp-2 cell slides and then subsequently stained and interpreted by two experienced observers independently. In parallel, ANA tests were also performed using a semi-automated indirect immunofluorescence assay (IFA) benchtop analyzer (Akiron NEO) and software using a standard kit as per the manufacturer's instructions. Positive/negative results, staining patterns, and intensities were compared between in-house HEp-2 cell slides as a reference standard and the semi-automated assay. Relative sensitivities, specificities, and degree of concordance were assessed by overall percentage agreement or by Cohen’s k.

Results: Among a total of 36 samples, in-house HEp-2 cell slides yielded 26 (72.22%) positive and 10 (27.78%) negative. The semi-automated IFA system yielded 21(58.33%) positive samples and 15(41.66%) negative samples. Overall relative sensitivity, specificity, and concordance of the semi-automated IFA system were 65.38%, 60%, and 63.88% (k = 0.21), respectively (p = 0.26).

Conclusions: The semi-automated IFA system (Akiron NEO) has fair agreement with in-house HEp-2 cell slides with respect to positivity/negativity and pattern recognition but poor agreement for intensity. This makes semi-automated evaluation of ANA relatively less suitable for routine testing.

Downloads

Download data is not yet available.

References

1. Choi HW, Kwon YJ, Park JH, Lee SY, Chun S, Won EJ, et al. Evaluation of a fully automated antinuclear antibody indirect immunofluorescence assay in routine use. Front Immunol 2020; 11: 607541. https://doi.org/10.3389/fimmu.2020.607541

2. Pashnina IA, Krivolapova IM, Fedotkina TV, Ryabkova VA, Chereshneva MV, Churilov LP, et al. Antinuclear autoantibodies in health: autoimmunity is not a synonym of autoimmune disease. Antibodies 2021; 10(1): 9.

https://doi.org/10.3390/antib10010009

3. Damoiseaux J, Andrade LE, Carballo OG, Conrad K, Francescantonio PL, Fritzler MJ, et al. Clinical relevance of HEp-2 indirect immunofluorescent patterns: the International Consensus on ANA patterns (ICAP) perspective. Ann Rheum Dis 2019; 78(7): 879-889.

https://doi.org/10.1136/annrheumdis-2018-214436

4. Mazzone R, Zwergel C, Artico M, Taurone S, Ralli M, Greco A, et al. The emerging role of epigenetics in human autoimmune disorders. Clin Epigenetics 2019; 11(1): 34.

https://doi.org/10.1186/s13148-019-0632-2

5. Irure-Ventura J, López-Hoyos M. The past, present, and future in antinuclear antibodies (ANA). Diagnostics 2022; 12(3): 647.

https://doi.org/10.3390/diagnostics12030647

6. Andrade LEC, Damoiseaux J, Vergani D, Fritzler MJ. Antinuclear antibodies (ANA) as a criterion for classification and diagnosis of systemic autoimmune diseases. J Transl Autoimmun 2022; 5: 100145.

https://doi.org/10.1016/j.jtauto.2022.100145

7. Infantino M, Meacci F, Grossi V, Manfredi M, Benucci M, Merone M, et al. The burden of the variability introduced by the HEp-2 assay kit and the CAD system in ANA indirect immunofluorescence test. Immunol Res 2017; 65(1): 345-354.

https://doi.org/10.1007/s12026-016-8845-3

8. Tebo AE. Recent approaches to optimize laboratory assessment of antinuclear antibodies. Clin Vaccine Immunol 2017; 24(12): e00270-17.

https://doi.org/10.1128/CVI.00270-17

9. Rojas G. Understanding and modulating antibody fine specificity: lessons from combinatorial biology. Antibodies 2022; 11(3): 48.

https://doi.org/10.3390/antib11030048

10. Ostrov BE. Reliability and reproducibility of antinuclear antibody testing in pediatric rheumatology practice. Front Med 2022; 9: 1071115.

https://doi.org/10.3389/fmed.2022.1071115

11. Chan EK, Damoiseaux J, de Melo Cruvinel W, Carballo OG, Conrad K, Francescantonio PL, et al. Report on the second International Consensus on ANA Pattern (ICAP) workshop in Dresden 2015. Lupus 2016; 25(8): 797-804. https://doi.org/10.1177/0961203316640920

12. Tipu HN, Bashir MM. Determination of specificity and pattern of antinuclear antibodies (ANA) in systemic rheumatic disease patients positive for ANA testing. J Coll Physicians Surg Pak 2018; 28(1): 40-43. https://doi.org/10.29271/jcpsp.2018.01.40

13. Arshad MZ, Ahmad D, Hussain M, Tipu HN, Riaz MO, Tanveer A. Development and validation of in-house HEp-2 cell slides for detection of antinuclear antibodies (ANA) by indirect immunofluorescence. J Coll Physicians Surg Pak. 2023; 33(3): 292-296.

https://doi.org/10.29271/jcpsp.2023.03.292

14. Van Beers JJBC, Hahn M, Fraune J, Mallet K, Krause C, Hormann W, et al. Performance analysis of automated evaluation of antinuclear antibody indirect immunofluorescent tests in a routine setting. Autoimmun Highlights 2018; 9(1): 8.

https://doi.org/10.1007/s13317-018-0108-y

15. Li Z, Han R, Yan Z, Li L, Feng Z. Antinuclear antibodies detection: a comparative study between automated recognition and conventional visual interpretation. J Clin Lab Anal 2019; 33(1): e22619.

https://doi.org/10.1002/jcla.22619

16. Sowa M, Hiemann R, Schierack P, Reinhold D, Roggenbuck D, Conrad K. Next-generation autoantibody testing by combination of screening and confirmation—the CytoBead technology. Clin Rev Allergy Immunol 2017; 53(1): 87-104.

https://doi.org/10.1007/s12016-016-8574-3

17. Cinquanta L, Bizzaro N, Pesce G. Standardization and quality assessment under the perspective of automated computer-assisted HEp-2 immunofluorescence assay systems. Front Immunol 2021; 12: 638863.

https://doi.org/10.3389/fimmu.2021.638863

18. Ricchiuti V, Adams J, Hardy DJ, Katayev A, Fleming JK. Automated processing and evaluation of anti-nuclear antibody indirect immunofluorescence testing. Front Immunol 2018; 9: 927. https://doi.org/10.3389/fimmu.2018.00927

Downloads

Published

30-10-2026

Issue

Section

Original Articles

How to Cite

1.
Arshad MZ, Riaz MO, Hussain M, Ahmad D, Tanveer A, Hassan MA. Evaluation Of A Semi-Automated Indirect Immunofluorescence Assay Using Validated In-House Hep-2 Cell Slides As A Reference Standard For Antinuclear Antibody Detection. Pak Armed Forces Med J [Internet]. 2026 Oct. 30 [cited 2026 Oct. 30];76(6):775-9. Available from: https://www.pafmj.org/PAFMJ/article/view/10690